CM30 Series Coating Thickness Gauge
Model:CM30F / CM30FH / CM30N / CM30FN
Measuring Range
CM30F :0-3000um
CM30FH:0-1000um
CM30N :0-2000um
CM30FN:0-1500um
Model:CM30F / CM30FH / CM30N / CM30FN
Measuring Range
CM30F :0-3000um
CM30FH:0-1000um
CM30N :0-2000um
CM30FN:0-1500um
Principle of Measurement
Magnetic Principle(Fe):Measure the thickness of non-magnetic coatings (such as paint layer, various anticorrosive coating, coating, powder spraying, plastic, rubber, synthetic materials, phosphating layer, chromium, zinc, lead, aluminum, tin, cadmium, etc.) on ferromagnetic metal substrates.
Eddy Current Principle (NFe) : Measure the thickness of all non-conductive coatings (such as rot coating, coating, powder coating, plastic, rubber, synthetic materials, oxide film, phosphating film) on non-ferromagnetic substrates like copper, aluminum and stainless steel.
Display | 2.4'' (320×240) IPS |
Working Principle | Magnetic / Eddy Current |
Unit | Metric(μm/mm) / Imperial(mils) |
Resolution | High / Low (metric mode only) |
Language | Chinese / English |
Statistic Data | Number of Measured Values, Average Value, Maximum Value, Minimum Value, Standard Deviation, Variable Coefficient |
Calibration Method | Zero Calibration, One-Point Calibration, Two-Point Calibration |
Number of Files | 200 |
Storage Capacity | 200,000 |
Backlight | Adjustable 6 levels |
Communication | Communication Software (USB data transfer) |
Power Off | Auto power off after 3 min. no operation or manual off |
Display Mode | Large Font Mode, Statistics Mode and Trend Graph Mode |
Protection level | IP68 |
Power Supply | 2pcs 1.5V AA Batteries |
Operating Time | 20 Hours |
Working Environment | -10 to +50℃ no strong magnetic field |
Size | 121.5mm*63.5mm*31.5mm |
Weight | 317g (No probe and no battery) |
Guarantee | 1 Year |
Instrument Model | CM30F | CM30N | CM30FH | CM30FN | |
Probe Type | F3 | N2 | F10 | FN1.5 | |
Measuring Principle | Magnetic | Eddy Current | Magnetic | Magnetic / Eddy Current | |
Measuring range (mm) | 0~3000 | 0~2000 | 0~10000 | 0~1500 | |
Low Range Resolution(mm) | 0.1 | 0.1 | 0.1 | 0.1 | |
Tolerance(mm) | ±(2%H+2) | ±(2%H+2) | ±(2%H+10) | ±(2%H+2) | |
Testing Condition (mm) | Min. Curvature Radius(convex) | 5 | 5 | 10 | 5 |
Min.Measuring Diameter | F20 | F20 | F40 | F20 | |
Min. Substrate Thickness | 0.5 | 0.5 | 2 | 0.5 |